DESIGNING AND MANUFACTURING TRITIUM MANAGEMENT TECHNOLOGY.
DESIGNING AND MANUFACTURING TRITIUM MANAGEMENT TECHNOLOGY.
Tyne Engineering's Wall Mounted Room Tritium In Air Monitor: Model 7065 is intended for room monitoring for both tritium and gamma with the convenience of simple wall mounting for ease of removal for calibration or relocation. The 7065 instrument is based on the same platform used in Tyne’s successful Portable Tritium-In-Air Monitor: Model 7043.
The wall mount version employs four ion chambers that are approximately 3 times larger than those used in their Model 7043 hand held instrument, hence increasing sensitivity by a factor of 3. The wall mounted instrument also employs a larger and more powerful, low noise, and high reliability pump, which is easily accessible by removal of the front panel. This room tritium monitor also allows for easy removal of the wireless transmitter for independent tune-up of this unit.
Compensation for gamma background is crucial in tritium measurement because a gamma field of 1mR/Hr will generate 500 times the ionization generated by l uCi/m3 of tritium. Tyne designed the four ion chambers in a symmetrical arrangement with each chamber connected to an ultra sensitive electrometer amplifier and filter. All signals pass through an ADC converter and a micro-processor calculates and displays the measured tritium values. The micro-processor also monitors air flow rate using a solid-state flow meter and controls the chamber temperatures during the heated purge cycle. A separate small ion chamber is used to cover higher tritium concentration ranges.
The front mounted LCD screen is a touch sensitive screen used to control device settings such as alarm levels. The instrument has both visual and audible alarm annunciation. The audible alarm can be canceled without canceling the visual annunciation. The touch screen panel is accessed by a two level password system. One password provides high level access allowing exclusive personnel to set up the device and the other provides restricted access enabling operators to only adjust certain features. The touch screen is augmented by use of a large digital read-out providing better, longer distance, visibility for operators.
The unit employs a front mounted high efficiency inlet air filter to ensure that the ion chambers do not become contaminated. If required, a sniffer probe adapter can be provided. The sniffer allows operators to locate suspected tritium leaks in the facility. For example, sniffing around a seal or flange to identify leaks.
Instrument output includes external communication via RS232, Ethernet and industry standard wireless to enable downloading or recording of information on a computer. The thin-film transistor (TFT) color LCD display mounted on the surface plate clearly displays instrument readings in both digital and graphic format showing real time measurements and trends.
The large ion chambers, configured in a compact internal module, enable a sensitivity that precludes the need for Methane or P-10 counting gases and the use of less reliable flow proportional counting systems. The ion chambers are robust and reliable, but if they become contaminated they can be purged by using an internally mounted heater while simultaneously pumping dry air through the chambers.
This wall mounted unit is powered by internal power supplies and only needs a standard 120 VAC three prong outlet to operate.
SPECIFICATIONS
Measuring chamber effective volume 1.5L,(2X0.75L)
Pump Positive displacement gas pump: 12 VDC, flow rate up to 8 L/min, discharge pressure up to 30 psig, up to 90% relative vacuum
Tritium measuring range 0.2μCi/m3 (7,400Bq/ m3) to 1Ci/m3(3.7E10 Bq/m3)
Ion Chamber Configuration Cruciform configuration: two for measuring and two for compensation
Radiation measured Total tritium and total gamma
Sensitivity 0.2μCi/m3 (0.01 DAC)
Accuracy 0.4μCi/m3 from 0 μCi/m3 to 100μCi/ m3; ±10% from 100μCi/ m3 to 1Ci/ m3
Zero stability ±0.4μCi/ m3
Gamma measurement Gamma is measured simultaneously with tritium. Measuring range is from 0.1mR/ h to 10 R/hr
Radon compensation Instrument provides radon compensation
Gamma background compensation less than ±10% or 5μCi/m3 reading change whichever is greater response to 20mR/Hr
Ion trap Built into each ion chamber
Flow rate Up to 8 L/min
Response time 2 to 5s with 0-15 m air hose, within 30s with hose from 100 to 150m at 90% of signal
Drift ±3% per 24 hrs; ±10% per 30 days
Alarms Local audio and visual alarms for tritium, gamma, air flow, detector failure and micro processor failure. Remote data and alarm display via RS-232 or analog output (0 to 3 VDC)
Alarm Response < 30S with 2-150m hose
Ion Chamber purge and decontamination Chambers can be heated using an internal temperature controlled cartridge heater
Condensation Prevention The heater cartridge temperature can be raised above ambient temperature to prevent condensation in instrument
Communication port RS232 (short distance)
Data Communication Ethernet connection provided Industrial wireless connectivity optional
per customer specification
Data Logging Frequency The tritium and gamma readings can be logged at a defined interval. Each data reading is date and time stamped
Displays One 2.8” TFT LCD touch screen display showing the values of tritium and gamma radiation in both numerical and graphical format. One LED screen 6 x 3⁄4” digits showing tritium concentration
Power supply 120 VAC (±10%) Standard; 220VAC (±10%) upon request; Power consumption < 3A
Temperature Operating Range 0 to 50°C degrees
Humidity Operating range 0-90%, Non-condensing
Size Approx 13.25" High x 8.1” Deep x 18.25" Wide
Weight Less than 30lbs with a steel case; Less than 22lbs with an aluminum case
Mounting Unit is designed for wall mounting on a pre-installed wall plate. Unit hangs from two upper hooks and is retained by a single screw at the base for positive and secure installation
Tyne Engineering has designed and manufactured a wide range of tritium air monitors. Our products can be incorporated into customized tritium management system design requirements. Applications include:
Tyne offers expedited delivery of two of their popular surface activity monitors:
Tyne Engineering has developed instruments specifically for measuring tritium inside process lines. Applications include:
We are here to ensure our customers receive industry leading products and service.
Copyright © 2023 tyne usa - All Rights Reserved.
Phone: 1.877.969.TYNE or 1.877.969.8963 Fax: 1.877.715.5508 Email: sales@tyneusa.com